Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
830User Manual 1178.8520.02 ─ 01
Manual operation: See " RF Attenuation Mode " on page 258
[SENSe:]LIST:RANGe<ri>:INPut:GAIN:STATe <State>
This command turns the preamplifier for a spurious emission measurement range on
and off.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See " Preamp " on page 258
[SENSe:]LIST:RANGe<ri>:INPut:GAIN[:VALue]
<Gain>
This command selects the preamplification level for the range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
[SENSe:]LIST:RANGe<ri>:LIMit:STARt <Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Level> Absolute limit at the start frequency of a SEM range.
Range: -400 to 400
*RST: 13
Default unit: dBm
Manual operation: See " Abs Limit Start / Abs Limit Stop " on page 259
[SENSe:]LIST:RANGe<ri>:LIMit:STATe <State>
This command turns the limit check for all spurious emission measurement ranges on
and off.
Suffix:
<ri>
.
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See " Limit Check " on page 259
Configuring and Performing Measurements

Table of Contents

Other manuals for R&S FSV3030

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSV3030 and is the answer not in the manual?

R&S FSV3030 Specifications

General IconGeneral
BrandR&S
ModelFSV3030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals