Rapid Shape Operating Instructions
(2) Carefully remove the build platform by
slowly pulling the platform out towards you
Caution: When removing, use the specially
designed removal tool to prevent dirtying
the machine and/or laboratory.
(3) Prevent the component thus removed from
being affected by UV and clean it directly
after removal
Prolonged UV radiation can polymerize any excess material remaining
on the component and thus impair the quality of the component.
7.5 Cleaning the component
General instructions:
In any case, consult the manufacturer’s material specifications when post-processing
(cleaning, if necessary, drying and post-exposure). A bio-compatibility can be achieved
only under consideration of the manufacturer’s specifications.
For cleaning the component, the Rapid Shape Part Cleaner or high-proof isopropanol
should be used. Under no circumstances, additives such as acetone or similar may be
used. These will damage the component immediately and will leave white marks.
(1) Put your components in the cleaning unit RS Wash.
(2) Ensure, that the isopropanol to be used is compatible with the material to be
washed.
(3) Select your print job or the used material.
(4) Start the washing process.
(5) After washing, remove the cleaned component out of the cleaning unit.
For more information about the handling of the cleaning unit RS wash, please refer to the
separately available operating instructions.