REA VERIFIER
© REA Elektronik GmbH, 64367 Mühltal, Germany --- REA MLV-2D Manual Version 3 – 21/12/2015
Page 61 of 98
Example of a verification report (long text format)
TransWin32, Version 1.0.16.1/11490r
F. Verweyen
VRF, REA Elektronik GmbH
Teichwiesenstrasse 1, Mühltal-Waschenbach Deutschland-Germany
Comment: Manual Documentation
Operator: Verweyen
Job number: Test job 4711
Aperture: 14 mil, 0.356 mm
Light source: 660nm / 45° Q
Lens: Pentax 35 mm
Unit: Micrometer [µm]
Date: 02/22/2012 10:22:19 AM
Device: REA MLV-2D (V.1.0.16.0/11492r)
Serial number: 0030.025.100-00532-9
Verifier: Software V.1.0.16.0/11497r
Reflectance adjustment: 02/18/2012, Valid until 03/19/2012
Level Correction: 02/18/2012, Valid until 03/19/2012
Scale adjustment: 02/18/2012, Valid until 03/19/2012
Symbology: GS1 Data Matrix ECC200
Code content (HR) (01)01234567890123(21)12345678901234
(17)050309(10)A123456x
Data structure (GS1)
1) 01 AI
2) 0123456789012 GTIN
3) 3 Check digit
Invalid Check digit in field,actual : 3, nominal : 8
3) 21 AI
4) 12345678901234 SERIAL
5) [GS] Field separator
6) 17 AI
7) 050309 USE BY / EXPIRY
8) 10 AI
9) A123456x BATCH/LOT
Data structure: Fail 0
User comment:
This is a test comment
Evaluation: ISO/IEC 15415, ISO/IEC 16022, GS1 General Specifications
Symbol: Fail 0.0/14/625/45° Q
Error list:
01) Data structure analysis failed
02) The modulation value is too low