Chapter 8
8-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B
Component Removal/Installation
8.1 Disconnecting
Power
Before removing any components from the ventilator, disconnect AC and
battery power sources (Figure 8-1).
1. Turn the ventilator OFF (O) and disconnect the AC power cord.
2. Remove the 2 battery cover screws.
3. Remove the battery cover and battery pack.
4. Disconnect the battery
Figure 8-1: Disconnecting Battery Power
Battery cable
connector
Battery compartment
Screws (x2)
(5 lbf-in., 57 N cm)