Chapter 2
Warnings and Cautions
REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc. 2-3
2.2 Cautions Federal law (US) restricts this device to sale by, or on the order of,
a physician.
Failure to replace a dirty filter may cause the device to operate at
higher than normal temperatures and damage the device.
Use only Respironics-approved filters.
Use only Respironics-approved batteries.
The gray reusable air inlet filter must be completely dry before
use. Never place a wet filter into the device.
Never install more than one gray reusable air inlet filter into the
ventilator. The ventilator is designed to accommodate one gray
reusable air inlet filter and one white disposable ultra-fine filter.
During battery charging, disconnecting AC mains power before
charging completes may result in a Battery Depleted or Missing
alarm. If a Battery Depleted or Missing alarm occurs, connect AC
power immediately to avoid total loss of power.
To prevent the premature depletion and reduced life expectancy of
the battery, store the ventilator with AC power connected and the
ON/OFF switch in the ON (I) position. When the ventilator is
stored disconnected from AC power with the power ON/OFF switch
ON ( I ) and in standby, the battery may become depleted within
two weeks. When the ventilator is stored with the ON/OFF switch
in the OFF (O) position, the battery typically becomes depleted in
three months.
If a 343 error code is displayed, remove the battery and charge it
using the BiPAP Focus battery charging cable (P/N 1040421).
This cable provides the initial charge required so that the
ventilator can recharge a depleted battery. Refer to the battery
charge instructions supplied with the cable.
Replace any filter that is damaged.
Do not attempt to clean or reuse disposable air inlet filters.
Electronic components used in this device are subject to damage
from static electricity. Perform repairs only in an antistatic,
electrostatic discharge (ESD)-protected environment.
Avoid exposing the ventilator to extreme environmental conditions.
Allow the ventilator to return to room temperature before
operation.
Never place liquids on or near the BiPAP Focus.
The information in this manual is provided for service personnel
reference.
Do not pull or crimp cables, tubes, or wires.