Chapter 1
Introduction and Intended Use
REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc. 1-3
1.2 Where to Go for
Help
For customer service and product support contact:
Respironics, Inc.
USA telephone: 1-800-345-6443
Fax: 1-800-886-0245
International telephone: 724-387-4000
International fax: 724-387-5012
service@respironics.com.
clinical@respironics.com.
www.respironics.com.
Pliers Local supplier
Pliers, needle nose Local supplier
Screwdriver, flat, #2 Local supplier
Screwdriver, flat, pen size Local supplier
Screwdriver, Phillips, #1 Local supplier
Screwdriver, Phillips, #2 Local supplier
Screwdriver, Phillips, pen size Local supplier
Torque driver (1 to 25 in.-lbs) Local supplier
Tubing, silicone, 3/16-in. ID x 6.5 ft. Local supplier
Table 1-2: Measurement Accuracy Requirements
Unit of
Measure
Range Accuracy
Pressure -25 to 150 cmH20
0-100 psi
+/- 1% of reading or +/- 0.20 cmH20 (whichever is
greater).
Flow (standard) 0-300 SLPM +/- 2% of reading +/- 0.20 SLPM (whichever is greater)
Torque 5 to 25 in-lbs. +/- 1 in-lbs. of reading
Voltage DC: +/- 5 to 50V
AC: 2 to 300 V
+/- 2% of reading
Table 1-1: Test Equipment, Tools, and Materials (Continued)
Description Manufacturer and Model