RIGOL DG1000 Command System
Programming Guide for DG1000
2-38
Detailed information of each command:
1.
PM:SOURce
Syntax PM:SOURce {INTernal|EXTernal}
Function Select internal or external modulation source, the default is INT.
Example PM:SOUR EXT
2.
PM:SOURce?
Syntax
PM:SOURce?
Function
Query the modulation source of PM.
Return Value The query returns INT or EXT.
3.
PM:INTernal:FUNCtion
Syntax PM:INTernal:FUNCtion
{SINusoid|SQUare|RAMP|NRAMp|TRIangle|NOISe|USER}
Function
Select the internal modulating wave of PM.
Explanation In internal modulation mode, the modulating
wave could be sine,
square, ramp, ne
gative ramp, triangle, noise or arbitrary wave, the
default is sine.
Example PM:INT:FUNC SQU
4.
PM:INTernal:FUNCtion?
Syntax
PM:INTernal:FUNCtion?
Function
Query the internal modulating wave selected .
Return Value The query returns SIN, SQU, RAMP, NRAM, TRI, NOIS or USER.
5.
PM:INTernal:FREQuency
Syntax PM:INTernal:FREQuency {<frequency>|MINimum|MAXimum}
Function Set the frequency of PM internal modulation in Hz.
Explanation Frequency range: 2mHz to 20kHz
Example PM:INT:FREQ 200
6.
PM:INTernal:FREQuency?
Syntax
PM:INTernal:FREQuency?
Function
Query the frequency of PM internal modulation.
Return Value The query returns the frequency of PM internal modulation in
scientific notation and the default unit is Hz, such as: