Chapter 2 Command System RIGOL
MSO7000/DS7000 Programming Guide 2-205
:TRIGger:CAN (Option)
Command List:
:TRIGger:CAN:BAUD
:TRIGger:CAN:SOURce
:TRIGger:CAN:STYPe
:TRIGger:CAN:WHEN
:TRIGger:CAN:SPOint
:TRIGger:CAN:LEVel
:TRIGger:CAN:BAUD
Syntax
:TRIGger:CAN:BAUD <baud>
:TRIGger:CAN:BAUD?
Description
Sets or queries the signal rate of CAN trigger. The unit is bps.
Parameter
Name
Type
Range
Default
<baud>
Integer
10 kbps to 5 Mbps
1 Mbps
Remarks
If the baud rate is set to a value with "M", then "A" should be added at the end of the
value. For example, if you send 5 M, you need to send 5 MA.
Return
Format
The query returns an integer ranging from 10 kbps to 5 Mbps.
Example
:TRIGger:CAN:BAUD 125000 /*Sets the signal rate to 125000 bps.*/
:TRIGger:CAN:BAUD? /*The query returns 125000.*/
:TRIGger:CAN:SOURce
Syntax
:TRIGger:CAN:SOURce <source>
:TRIGger:CAN:SOURce?
Description
Sets or queries the trigger source of CAN trigger.
Parameter
Name
Type
Range
Default
<source>
Discrete
{D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D1
1|D12|D13|D14|D15|CHANnel1|CHANnel2|C
HANnel3|CHANnel4}
CHANnel1
Return
Format
The query returns D0, D1, D2, D3, D4, D5, D6, D7, D8, D9, D10, D11, D12, D13, D14,
D15, CHAN1, CHAN2, CHAN3, or CHAN4.
Example
:TRIGger:CAN:SOURce CHANnel2 /*Sets the trigger source to CH2.*/
:TRIGger:CAN:SOURce? /*The query returns CHAN2.*/

Table of Contents

Other manuals for Rigol MSO7000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol MSO7000 Series and is the answer not in the manual?

Rigol MSO7000 Series Specifications

General IconGeneral
BrandRigol
ModelMSO7000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals