EasyManua.ls Logo

Rigol MSO7000 Series

Rigol MSO7000 Series
259 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 Command System RIGOL
MSO7000/DS7000 Programming Guide 2-47
Example
:COUNter:SOURce CHANnel2 /*Sets the source of the frequency
counter to CH2.*/
:COUNter:SOURce? /*The query returns CHAN2.*/
:COUNter:MODE
Syntax
:COUNter:MODE <mode>
:COUNter:MODE?
Description
Sets or queries the mode of the frequency counter.
Parameter
Name
Type
Range
Default
<mode>
Discrete
{FREQuency|PERiod|TOTalize}
FREQuency
Return
Format
The query returns FREQ, PER, or TOT.
Example
:COUNter:MODE PERiod /*Sets the mode of the frequency counter to
PERiod.*/
:COUNter:MODE? /*The query returns PER.*/
:COUNter:NDIGits
Syntax
:COUNter:NDIGits <val>
:COUNter:NDIGits?
Description
Sets or queries the resolution of the frequency counter.
Parameter
Name
Type
Range
Default
<val>
Integer
3-6
5
Remarks
The resolution setting is only available for "Period" and "Frequency", and
unavailable for "Totalize".
Return
Format
The query returns an integer ranging from 3 to 6.
Example
:COUNter:NDIGits 4 /*Sets the resolution of the frequency
counter to 4.*/
:COUNter:NDIGits? /*The query returns 4.*/
:COUNter:TOTalize:ENABle
Syntax
:COUNter:TOTalize:ENABle <bool>
:COUNter:TOTalize:ENABle?
Description
Enables or disables the statistical function of the frequency counter; or
queries the on/off status of the statistical function of the frequency counter.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{1|ON}|{0|OFF}}
0|OFF
Remarks
Not available for "Period" and "Frequency".
Return
Format
The query returns 1 or 0.