Page #227 background imageLoading...
Page #227 background image
Chapter 2 Command System RIGOL
MSO7000/DS7000 Programming Guide 2-213
:TRIGger:FLEXray (Option)
Command List:
:TRIGger:FLEXray:BAUD
:TRIGger:FLEXray:LEVel
:TRIGger:FLEXray:SOURce
:TRIGger:FLEXray:WHEN
:TRIGger:FLEXray:BAUD
Syntax
:TRIGger:FLEXray:BAUD <baud>
:TRIGger:FLEXray:BAUD?
Description
Sets or queries the signal rate of FlexRay trigger. The default unit is bps.
Parameter
Name
Type
Range
Default
<baud>
Discrete
{2500000|5000000|10000000}
10000000 bps
Return
Format
The query returns 2500000, 5000000, or 10000000.
Example
:TRIGger:FLEXray:BAUD 5000000 /*Sets the signal rate to 5000000 bps.*/
:TRIGger:FLEXray:BAUD? /*The query returns 5000000.*/
:TRIGger:FLEXray:LEVel
Syntax
:TRIGger:FLEXray:LEVel <level>
:TRIGger:FLEXray:LEVel?
Description
Sets or queries the trigger level of FlexRay trigger. Its unit is the same as that of the
current amplitude.
Parameter
Name
Type
Range
Default
<level>
Real
(-5 × VerticalScale - OFFSet) to
(5 × VerticalScale - OFFSet)
0 V
Remarks
For VerticalScale, refer to the :CHANnel<n>:SCALe command; for OFFSet, refer to
the :CHANnel<n>:OFFSet command.
Return
Format
The query returns the trigger level in scientific notation.
Example
:TRIGger:FLEXray:LEVel 0.16 /*Sets the trigger level to 160 mV.*/
:TRIGger:FLEXray:LEVel? /*The query returns 1.600000E-1.*/

Table of Contents

Other manuals for Rigol MSO7000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol MSO7000 Series and is the answer not in the manual?

Rigol MSO7000 Series Specifications

General IconGeneral
BrandRigol
ModelMSO7000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals