Rishabh RISH CON - I - 3.1.3 Communication Parameter Setting

Rishabh RISH CON - I
65 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Secondary value edit” mode.
Pressing the “ Up” key will confirm the CT Secondary &
advance to the Communication parameter Setting (section
3.1.3).
3.1.3. Communication Parameter Setting
3.1.3.1 Address Setting
This screen applies to the RS 485 output only.
This screen allows the user to set RS485 parameter for
instruments. The range of allowable address is 1 to 247
Pressing Down” key will
advance to the “New/Change
address value edit” mode.
Pressing the Up” key will
confirm the present value as
Address and advance to Baud
Rate selection (3.1.3.2).
New/Change Address value
(*Denotes that digit will be flashing).
First digit is always blank.
Pressing the “ Down “ key will
scroll the value of the second digit
from 0 through to 2, value will wrap
from 2 round to 0.
Pressing the Up” key will
advance the operation to the next
digit and set the second digit, in this case to “0”
Enter New/change Address value, second digit entered,
prompting for third digit. (*Denotes that digit will be flashing).
Pressing the Down” key will scroll the value of the third digit
13