GX-3R Operator’s Manual Chapter 3: Operation 31
Responding to System Failure Alarms
1 . If a system failure occurs, the system failure screen displays an error code as shown
below:
2 . The error code meanings are shown in the table below:
3 . If the error code is anything but 031 as shown above, the instrument cannot be used.
Contact RKI Instruments, Inc. as soon as possible.
If the error code is 031, you may press and release POWER MODE to continue into
Measuring Mode if the instrument must be used temporarily.
CAUTION: There will be no datalogging function if you operate the instrument after a 031
system failure. Contact RKI Instruments, Inc. as soon as possible.
Table 6: Error Code Explanation
Error Code Explanation
000 ROM failure
010 RAM failure
021 FRAM failure
031 FLASH memory failure
080 Acceleration sensor failure
081 PCB failure
082 Temperature sensor failure
SYSTEM
7:49
031
FAIL