58 • Chapter 5: User Mode and Calibration GX-3R Operators Manual
Setting the Calibration Values in CAL--P
1 . While in User Mode, press AIR to scroll to GAS CAL.
2 . Press and release POWER MODE. The AIR CAL menu item appears.
3 . Use AIR to scroll to the AUTO CAL menu item.
4 . Press and release POWER MODE. The Auto Cal screen for gases assigned to Cylinder
A appears.
5 . Use AIR to scroll to CAL--P.
6 . Press and release POWER MODE. The combustible gas channel appears.
GAS CAL
M 7:49
AUTO CAL
M 7:49
CAL--P
M 7:49
CAL--P
M 7:49
- - -
CH4 %LEL