4 Operation
Manual TME400-VMF · EN02 · 2018 September 6th
62
X12
Delete parameter ar-
chive (E)
0
No (default)
1
Yes
X13
Parameter archive (E)
fill level
Display value
Coor-
dinate
Name
Modbus
register
Modbus
access
Protec-
tion
Data
type
Min.
Max.
Default
Unit
X01
Time
712
W
E
string8
T
X02
Date
717
W
E
string8
D
X10
Delete parameter
archive
722
W
E
menu16
0
1
0
X11
Parameter archive
fill level
723
R
A
unit16
-
-
0
%
X14
Delete event
archive
726
W
E
menu16
0
1
0
X15
Event archive
fill level
727
R
A
unit16
-
-
0
%
X16
Measurement archive
mode
728
W
E
menu16
0
1
0
X17
Minute archive interval
729
W
E
menu16
0
2
0
X18
Delete minute archive
730
W
E
menu16
0
1
0
X19
Minute archive fill level
731
R
A
unit16
-
-
0
%
X20
Delete day archive
732
W
E
menu16
0
1
0
X21
Day archive fill level
733
R
A
unit16
-
-
0
%
X22
Delete month archive
734
W
E
menu16
0
1
0
X23
Month archive fill level
735
R
A
unit16
-
-
0
%
X12
Delete parameter
archive (E)
724
W
E
menu16
0
1
0
X13
Parameter archive (E)
fill level
725
R
A
unit16
-
-
0
%
4.3.3.7.
Settings
Coor-
dinate
Name
Description
Z04
X:Y maximum pulse
error
A differential circuit compares the metered pulse of measuring and
comparison channels alternatingly. Every deviation is counted inter-
nally. An alarm is generated if the adjusted limit value is exceeded.
The failure counter is reset to 0 for each new measurement or after
the maximum number of pulses (Z05) is reached.