4 Operation
MANUAL TME400-VCF · EN08 · July, 01
st
2021
69
Caution
Adoption of the correction changes the temperature charac-
teristic measured in the factory and stored. Only perform
this change if you are certain that you have detected a devi-
ating temperature measurement.
Of course, this correction is subject to the mandatory cali-
bration.
0
No (default)
1
Yes
Note
Correction values displayed in the coordinates D08 (f0) and
D09 (f1) are calculated internally. These correction values
may only change within the scope of 0.9 to 1.1; otherwise
there is a defect which must be corrected by RMG.
Coor-
dinate
Name
Modbus
register
Modbus
access
Protec-
tion
Data
type
Min.
Max.
Default
Unit
D01
Temperature
324
R
A
float
-
-
-
°C
D02
Temperature mode
587
W
E
menü16
0
1
0
D03
Temperature default
588
W
E
float
-40.0
80.0
10.0
T-V
D04
Temperature Minimum
590
W
E
float
-40.0
80.0
-25.0
°C
D05
Temperature Maximum
592
W
E
float
-40.0
80.0
60.0
°C
D06
Temperature moderation
594
E
float
0.1
1.0
1.0
T-D
D11
PT1000 resistance
602
R
A
float
-
-
-
Ohm
D12
PT1000 resistance
uncorr.
604
R
A
float
-
-
-
Ohm
D30
Temperature
(uncorrected)
606
R
A
float
-
-
-
°C
D35
Temperature setpoint 1
616
W
N
float
-40.0
80.0
-10.0
°C
D36
Temperature setpoint 2
618
W
N
float
-40.0
80.0
50.0
°C
D37
Temp. actual value 1
620
W
N
float
-40.0
80.0
-10.0
°C
D38
Temp. actual value 2
622
W
N
float
-40.0
80.0
50.0
°C
D41
Write temperature
corr.
628
W
E
menü16