Roche Diagnostics
Urisys 1100® · ≥5.7 · Operator's Manual · 9.0
18 4. Software
Send results For sending stored results to the serial interface. Options
are the same as for PRINT RESULTS. Results can be
sent as often as required.
u See:
8.7 Data transmission to a PC or host computer (41)
11.1 Serial Interface (54)
Clear memory Erases results from memory.
Modes Choice of Printer, Measure and interface (Interf.)
modes.
Printer Printer options are:
• On: The printer is switched on. Each result is printed
once.
• 2 Copies: In Normal Mode, each result is printed
twice.
Note: After serial readings (FastMode) and when the
printout is a repeat printout (activated by the PRINT
RESULTS function), each result is printed only once.
• Off: For switching off the printer when the printout of
results is only required at the end of a series of
readings (activated by the PRINT RESULTS function),
or when results are to be sent via the interface to a
personal computer or host only.
Measurement Choice of measuring mode:
• NormalMode: Read cycle 70 seconds, incubation
within the analyzer.
u 8.2 Performing a measurement in Normal Mode (for
Single Readings) (34)
• FastMode: Read cycle 30 seconds, incubation
outside the analyzer.
u 8.4 Fast Mode (for Serial Readings) (38)
Interface Choice of unidirectional (Unidir.), bidirectional (Bidir.) or
ASTM data transfer.
u See:
8.7 Data transmission to a PC or host computer (41)
11. Connecting to other devices (54)
Logdata Choice of sending the log files to the host PC or printing
out the last 10 logged data. (Only by supervisors in
Authentication mode)
Strip setting For selecting, units and range limit settings.
Units Options are:
• Conventional units, Conv. (mg/dL)
• SI units, SI (mmol/L)
• Arbitrary units, Arbitr. (1+, 2+, 3+, 4+)