4. Software 19
Either single units or combined units (Conv./Arb. or
SI/Arb.) can be selected.
The operator selects a unit in which the results are to be
stored, printed and/or transferred to a computer. After a
new unit has been selected, the repeat printout (activated
by PRINT RESULTS) and all following printouts and/or
data transfer (activated by SEND RESULTS) will be in the
newly chosen unit.
Range limits For changing the reflectance range limits. Changing the
range limits for the result intervals alters evaluation
sensitivity.
u See:
4.4 Results table (21)
4.5 Changing the range limits (22)
Language This enables the display language to be set. The Other
option allows the operator to choose between English,
German, Italian, Spanish and French.
Device ID Displays the 5 digits device ID, which is part of the factory
settings and cannot be changed. The device ID will also
be sent to the host.
Operator ID Choice between On and Auth. modes for the operator
identification. If activated the input of an operator ID will
be required by the start of the analyzer.
The Operator ID will appear in the result printout and will
also be sent to the host PC. Authentication mode offers a
lockout function and requires the ASTM protocol.
u See:
5.2 Authentication (28)
8.6 Entering Patient ID, Operator ID and
Authentication Password (39)
Parameter For selecting the order and number of parameters in the
patient report and for unidirectional (Unidir.) data
transfer.
To change the order or number of parameters in the
patient report, press the function key on the left (New)
with the display showing SEQUENCE PARAMETERS.
This calls up the first item in the patient report (1.
PARAM.: SG) followed by the first parameter. Press Ye s
to confirm the parameter in the position shown. Pressing
No causes the next test strip parameter to be offered for
the first position (display = 1. PARAM.: PH; etc.).