EasyManuals Logo

Roche cobas Urisys 1100 User Manual

Roche cobas Urisys 1100
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
Roche Diagnostics
Urisys 1100® · 5.7 · Operator's Manual · 9.0
36 8. Reading test strips
5
Press the START button.
I An acknowledging beep sounds.
6
After warming up time, the test strip tray advances
slightly, the retaining bar (B) closes, and the grey
reference pad on the test strip tray is read.
7
ATTENTION: Ensure that the retaining bar is
locked into place and that the test strip is in the
correct position. If the test strip is not correctly located
in the middle of the test strip tray, move it gently to the
side until it is properly aligned (B). Be careful not to
move the test strip tray.
8
Fifty-five seconds after the START button is pressed,
the first test pad is measured, followed by the others.
f After that, the test strip tray returns to the start
position and the retaining bar opens.
f The result will be printed out and the next sample
number will be displayed.
9
Remove and dispose of the test strip.
Wipe any urine residues from the test strip tray
with a lint-free cloth.
10
The next test strip can be dipped, wiped off, placed in
the test strip tray and read by pressing the START
button.
q For details of what must be observed when
working with patient identification numbers, see:
u 8.6 Entering Patient ID, Operator ID and Authentication
Password (39)
5
6
B
7
B
9

Other manuals for Roche cobas Urisys 1100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas Urisys 1100 and is the answer not in the manual?

Roche cobas Urisys 1100 Specifications

General IconGeneral
BrandRoche
Modelcobas Urisys 1100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals