E6 CHIP ERROR!
Cause
The program chip on the right of the analyzer underneath
the printer cover is missing, is not making contact, is
defective or contains an old software version.
Action 1. Switch off the Urisys 1100
®
system.
2. Insert the program chip and switch the analyzer on
again.
If E6 CHIP ERROR! appears again, the analyzer is
defective.
E7 MISSING TRAY!
Cause No test strip tray inserted or test strip tray not inserted far
enough to be engaged by the motor.
Action 1. Insert the test strip tray correctly.
2. Press the START button.
u Inserting the test strip tray (24)
E8 TRAY POSITION ERROR!
Cause
The positioning hole in the test strip tray is soiled or still
wet after cleaning; the retaining bar is open while the test
strip tray is advancing or the retaining bar mechanism is
fouled with urinary deposit and blocked.
u 9.2 Cleaning the test strip tray (43)