Urisys 1100® · 5.7 · Operator's Manual · 9.0
10. Error messages and troubleshooting 53
Cause The analyzer is older than 5 years and all troubleshooting
steps failed.
Action Replace the analyzer.
Cause The analyzer is less than 5 years old and all
troubleshooting steps failed.
Action Contact your local Roche Diagnostics representative for
support.
NO PRINTOUT
Cause PRINTER: OFF has been selected, or the analyzer is
operating in FastMode, or the printer/software is
defective, or the printer is out of paper.
u 8.4 Fast Mode (for Serial Readings) (38)
Action 1. Insert paper if needed.
2. Choose PRINTER: ON(1 Copy) to re-activate the
printer.
3. Request a patient report via the Print function.
- If this fails, activate the Linefeed function.
If there is still no response, the analyzer is defective.
THE ANALYZER WILL NOT READ EVEN THOUGH THE SEQUENCE NUMBER IS
DISPLAYED
Action If an AT/PC keyboard is connected, press the Escape key,
or switch the Urisys 1100
®
system off and back on again.

Other manuals for Roche cobas Urisys 1100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas Urisys 1100 and is the answer not in the manual?

Roche cobas Urisys 1100 Specifications

General IconGeneral
BrandRoche
Modelcobas Urisys 1100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals