52 10. Error messages and troubleshooting
Action 1. Repeat the measurement with a new analyzer-
compatible test strip.
- Follow the directions carefully and ensure the test
strip is correctly inserted
u 8.2 Performing a measurement in Normal Mode (for
Single Readings) (34)
1. Repeat calibration if necessary.
2. Repeat the measurement.
If the problem persists, perform a quality control
u 7. Quality control (QC) (32)
Cause Electromagnetic interference from other devices.
u 5. Installation (23)
Action Remove external sources of interference, if there are any.
CONTROL VALUES FALL OUTSIDE THE DESIGNATED RANGES
Cause 1. Test strip incorrectly positioned.
2. The wrong test strip, such as Combur10Test
®
test
strips, may have been used.
Action 1. Repeat calibration if necessary.
2. Repeat the measurement with a new analyzer-
compatible test strip.
- Follow the directions carefully and ensure the test
strip is correctly inserted.
u 8.2 Performing a measurement in Normal Mode (for
Single Readings) (34)
Cause The test strip tray is older than 18 months and the above
described troubleshooting steps failed.
Action 1. Replace the test strip tray by a spare tray.
2. Calibrate the analyzer using a new calibration strip.
3. Repeat the measurement with a new analyzer-
compatible test strip.
- Follow the directions carefully and ensure the test
strip is correctly inserted
u 8.2 Performing a measurement in Normal Mode (for
Single Readings) (34)
Cause Electromagnetic interference from other devices
u 5. Installation (23)
Action Remove external sources of interference, if there are any.