Roche Diagnostics
Urisys 1100® · 5.7 · Operator's Manual · 9.0
10. Error messages and troubleshooting 51
E17 INVALID PASSWORD!
Cause The entered password doesn’t match.
Action Enter a correct password.
E18 INVALID OP.ID!
Cause The entered operator ID is not valid.
Action Enter a valid operator ID.
E19 LIST DOWNLOAD FAILED!
Cause The new operator ID list download failed.
Action No action.
After 2 seconds next state starts with old list if there was.
E20 NO VALID LIST!
Cause There is no valid list at all in the device.
Action Try to download an operator ID list from the host or
continue without authenticated operator.
CHECK MEASUREMENT
Analyzer prints out software and chip version number and
3-digit numbers without naming the parameters.
Cause Service function is activated.
Action Press the Back function key to return to the main menu.
VALUES OBTAINED ARE IMPLAUSIBLE WHEN COMPARED WITH THOSE FROM VISUAL
EVALUATION
Cause 1. Test strip incorrectly positioned, uncharacteristic test
pad colors, proper incubation intervals not kept to
during serial measurements.
2. The wrong test strip, such as Combur10Test
®
test
strips, may have been used.

Other manuals for Roche cobas Urisys 1100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas Urisys 1100 and is the answer not in the manual?

Roche cobas Urisys 1100 Specifications

General IconGeneral
BrandRoche
Modelcobas Urisys 1100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals