64 Rockwell Automation Publication 856T-UM001B-EN-P - April 2021
Appendix C Device Parameters
Diagnostics Tab
Vibration Configuration
Vibration Warning limit 0x0143/323 Read/write 16,000 100…16,000
RecordT
(bit length = 48)
X-axis Vibration Warning
Limit
0x0143/323 0x0001/1 Read/write 16,000 100…16,000
IntegerT
(bit length = 16,
bit offset = 32)
Y-axis Vibration Warning
Limit
0x0143/323 0x0002/2 Read/write 16,000 100…16,000
IntegerT
(bit length = 16,
bit offset = 16)
Z-axis Vibration Warning
Limit
0x0143/323 0x0003/3 Read/write 16,000 100…16,000
IntegerT
(bit length = 16,
bit offset = 0)
Vibration Calibration
period
0x0144/324 Read/write 10 0…480
UIntegerT
(bit length = 16)
Table 15 - Parameter Tab (Continued)
Parameter Name
Index
(Hex/Dec)
Sub Index
(Hex/Dec)
Access Default Value Allowed Value
Data Type
(Length, Offset)
Table 16 - Diagnostics Tab
Parameter Name
Index
(Hex/Dec)
Sub Index
(Hex/Dec)
Access Default Value Allowed Value
Data Type
(Length, Offset)
Device Access Locks 0x000C/12
Read/write 0 0…3
RecordT
(bit length = 16)
Read/write 0
Bit 0 - ParameterAccessLock
(0=unlocked)
(1= locked)
BooleanT
(bit length = 1
bit offset = 0)
Read/write 0
Bit 1 - DataStorageLock
(0=unlocked)
(1= locked)
BooleanT
(bit length 1
bit offset=1)
Service Status
Device Status 0x0024/36 Read-only 0
0: Device is operating properly
1: Maintenance Required
2: Out of Specification
3: Functional Check
4: Failure
(see Appendix D on page 69
for details)
UIntegerT
(bit length = 8)
Power On self-test 0x0137/311 Read-only 0
0…12
0: Success
1: Firmware Image Corrupted
5: RAM test Failed
6: Secondary Processor Failed
7: IO Link PHY Failed
9: Accelerometer Communication
Failed
10: Serial Flash Integrity Failed
11: Bad Class Voltage
12: Bad Class B voltage
(see Power On Self Test (POST)
on
page 43 for details)
UIntegerT
(bit length = 16)
Invalid Configuration 0x0150/336 Read-only 0
0: Valid Configuration
1: Invalid Configuration
UIntegerT
(bit length = 1)
Operation Information
Operating Hours since
inception
0x0151/337 Read-only 0 0…4,294,967,296
UIntegerT
(bit length = 32)
Operating Hours since
power up
0x005B/91 Read-only 0 0…4,294,967,296
UIntegerT
(bit length = 32)

Table of Contents