Measurement Control  CMU 
1100.4903.12 3.18  E-9 
Table 3-5  Measurements in function group Audio (with option CMU-B41) 
Menu   Function 
Analyzer/Generator 
Generates a single-tone sinusoidal audio signal and measures the DC and AC voltage and 
the Total Harmonic Distortion and Noise of a single-tone audio signal.  
Multitone 
Generates a composite audio signal consisting of up to 20 individual fixed-frequency tones 
with configurable frequency and level. An audio signal containing the same tones can be 
analyzed in a single measurement and displayed in a bar chart. A limit check is provided for 
all results.  
 
A graphical overview of the menus is given at the end of this chapter. 
 
 
 
 
General Settings 
A number of settings can be made in several of the configuration menus assigned to the individual 
measurement groups. In combination, these settings define the scope of each measurement, i.e. the 
number of results acquired and the type of results displayed. The following brief overview is intended to 
avoid confusion of terms. 
 
Application 
 
Applications are different measurements belonging to the same measurement 
group. They effectively split up a measurement group into various related 
subgroups which can be configured separately.  
They are selected via the Application softkey in the measurement menus. 
 
Statistic Count / 
Statistics Cycle 
 
The statistic count is equal to the integer number of evaluation periods which 
form one statistics cycle. The definition of an evaluation period changes from 
one measurement to another:  
•  The evaluation period for Spectrum  measurements (function group RF) 
corresponds to the duration of a sweep.  
•  The evaluation period for Multitone measurements (function group Audio) 
corresponds to the time until the system has settled and a valid result is 
available.  
•  In all digital radio communication systems (GSM, TDMA, CDMA, ...), the 
information is transmitted in periodic timeslots of equal length. These slots 
provide natural evaluation periods for all measurements on digital network 
tests. See the relevant manuals for more specific information.  
Depending on the repetition mode (see below), a measurement may extend 
over one or several statistics cycles. The statistic count is set in the Control tab 
of the configuration popup-menus assigned to each measurement group. If this 
parameter is omitted, a statistics cycle always comprises just one evaluation 
period.  
 
Repetition Mode 
 
The  repetition mode defines how many statistics cycles are measured if the 
measurement is not stopped by a limit failure (see stop condition On Limit 
Failure below). Two modes are available for all measurements: 
Single Shot  The measurement is stopped after one statistics cycle 
Continuous   The measurement is continued until explicitly terminated by 
the user; the results are periodically updated 
A third repetition mode is available in remote control: