CMU RF Measurements
1100.4903.12 6.55 E-10
Results – Subsystem SPECtrum
The subsystem SPECtrum starts the Spectrum measurement and returns the results. The subsystem
corresponds to the various output elements in the graphical measurement menu Spectrum.
READ:ARRay:SPECtrum[:CURRent]? Spectrum Results
READ:ARRay:SPECtrum:AVERage?
READ:ARRay:SPECtrum:MAXimum?
READ:ARRay:SPECtrum:MINimum? Start single shot meas. and return results RUN
FETCh:ARRay:SPECtrum[:CURRent]?
FETCh:ARRay:SPECtrum:AVERage?
FETCh:ARRay:SPECtrum:MAXimum?
FETCh:ARRay:SPECtrum:MINimum? Read meas. results (unsynchronized) RUN
SAMPle:ARRay:SPECtrum[:CURRent]?
SAMPle:ARRay:SPECtrum:AVERage?
SAMPle:ARRay:SPECtrum:MAXimum?
SAMPle:ARRay:SPECtrum:MINimum? Read results (synchronized) RUN
Returned value
Parameter description Def. value Default unit FW vers.
–128.0 dBm to + 48.0 dBm
...
–128.0 dBm to + 48.0 dBm
1
st
power value
560
th
power value
NAN
NAN
dBm
dBm
V1.15
Command description
These commands are always queries. They return the results of the spectrum measurement at 560 equidistant
test points. The measurement range is defined by the commands [SENSe:]SPECtrum:FREQuency:CENTer
and [SENSe:]SPECtrum:FREQuency: SPAN or by [SENSe:]SPECtrum:FREQuency:STARt and
[SENSe:]SPECtrum: FREQuency:STOP.
Note: The number of test points may be reduced at very narrow measurement ranges.