Popup Menu Setup  CMU 
1100.4903.12 4.12  E-10 
Serial Interfaces (Setup – Comm.) 
The interface menu (Setup Comm.) defines the transmission parameters of the serial outputs COM 1 
and COM 2. 
 
SETUP
 
 
Fig. 4-8  Interface menu 
 
COM 1 
 
The COM 1 section defines the transmission parameters for the serial interface 
COM 1. 
If required, the transmission parameters must be set such as to comply with the 
parameters of the addressed device (PC etc.). The interface characteristics are 
explained in detail in Chapter 8, Hardware Interfaces. The individual parameters 
are presented in Table 4-1 below. 
The Application field indicates whether the serial port is used as a printer connector 
(Printer, if COM 1 was selected as printer port in the Print tab; see p. 4.9) or for 
data transfer to the controller (Transfer, default setting). This parameter can not be 
changed in the Comm. tab.  
 
 
Remote control 
SYSTem:COMMunicate:SERial1:APPLication 
SYSTem:COMMunicate:SERial1:TRANsmit:PACE... etc. 
 
Table 4-1  Transmission parameters of the serial interfaces 
Parameter Meaning  Value range 
Application Addressed device  Transfer | Printer (fixed setting) 
Baud Rate  Data transmission rate  110 |300 | 600 | 1200 | 2400 | 4800 | 9600 | 19200 | 
38400 | 57600 | 115200 
Data Bits  Number of data bits  7 |  8 
Stop Bits  Number of stop bits  1 | 2 
Parity  Number of parity bits  none | odd | even 
Protocol Transmission protocol  none | XonXoff | CtsRts