EasyManua.ls Logo

Rohde & Schwarz CMU 300 - Test Tones - Subsystem Multitone

Rohde & Schwarz CMU 300
418 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CMU Audio Generator and Analyzer (with Option CMU-B41)
1100.4903.12 6.81 E-10
DEFault:MULTitone:LIMit:LINE <Enable> Default Settings
<Enable>
Description of parameters Def. value Def. unit FW vers.
ON |
OFF
The parameters are set to their default values
Some or all parameters differ from the default values
ON
– V3.0
Description of command
If used as a setting command with the parameter ON this command sets all parameters of the subsystem (includ-
ing AF1Channel and AF2Channel) to their default values (the setting OFF results in an error message). If used
as a query the command returns whether all parameters are set to their default values (ON) or not (OFF).
Test Tones – Subsystem MULTitone:...TDEFinition
The subsystem MULTitone:...TDEFinition configures the audio test signal used for the Multitone meas-
urement. The subsystem corresponds to the Tone Def. tab of the popup menu Multitone Configuration.
CONFigure:MULTitone:AF1Channel:TDEFinition Test Tones
CONFigure:MULTitone:AF2Channel:TDEFinition
<Freq_1>, <Lev_1>, <Enable_1>, ... <Freq_20>, <Lev_20>, <Enable_20>
<Freq_nr>
Description of parameters Def. value Def. unit
10 Hz to 15999 Hz, Frequency of test tone <nr> See below Hz
<Lev_nr>
Description of parameters Def. value Def. unit
1.0 µV to 5.0 V, Level at test tone <nr> See below V
0
<Enable_nr>
Description of parameters Def. value Def. unit FW vers.
ON | OFF Switch on / off test tone <nr> See below V3.0
Description of command
This command enables and configures up to 20 test tones. The minimum frequency spacing between two tones
is 1 Hz. The sum of all test tones must not exceed the maximum AF generator level quoted in the data sheet.
The following default test tones are provided:
Tone <nr> Frequency/[Hz] Level/[V] Enable Tone <nr> Frequency/[Hz] Level/[V] Enable
1 300 0.01 ON 11 1700 0.01 ON
2 440 0.01 ON 12 1840 0.01 ON
3 580 0.01 ON 13 1980 0.01 ON
4 720 0.01 ON 14 2120 0.01 ON
5 860 0.01 ON 15 2260 0.01 ON
6 1004 0.01 ON 16 2400 0.01 ON
7 1140 0.01 ON 17 2540 0.01 ON
8 1280 0.01 ON 18 2680 0.01 ON
9 1420 0.01 ON 19 2820 0.01 ON
10 1560 0.01 ON 20 3000 0.01 ON
Note: If the level of all test tones is derived from a total level (TLEVel setting in the
CONFigure:MULTitone:AF1Channel:TDEFinition:MODE command), the individual level
settings are ignored. The <Frequency> and <Enable> parameter settings are still effec-
tive.