CMU RF Measurements 
1100.4903.12 6.49  E-10 
Results – Subsystem POWer...?  
The subsystem POWer...? starts the power measurement and returns the results. The subsystem cor-
responds to the graphical measurement menu Power. 
 
READ:ARRay:POWer[:CURRent]?   Power Results
READ:ARRay:POWer:AVERage? 
READ:ARRay:POWer:MAXimum? 
READ:ARRay:POWer:MINimum?  Start single shot meas. and return results  ⇒ RUN
FETCh:ARRay:POWer[:CURRent]? 
FETCh:ARRay:POWer:AVERage? 
FETCh:ARRay:POWer:MAXimum? 
FETCh:ARRay:POWer:MINimum?  Read meas. results (unsynchronized)  ⇒ RUN
SAMPle:ARRay:POWer[:CURRent]? 
SAMPle:ARRay:POWer:AVERage? 
SAMPle:ARRay:POWer:MAXimum? 
SAMPle:ARRay:POWer:MINimum?  Read results (synchronized)
  ⇒ RUN
Returned value 
Parameter description  Def. value  Default unit  FW vers. 
–128.0 dBm to + 48.0 dBm, 
..., 
–128.0 dBm to + 48.0 dBm 
1
st
 value for power 
 
500
th
 value for power 
NAN 
 
NAN 
dBm 
 
dBm 
V 1.15 
Command description 
These commands are always queries. They return the power values versus time at 1000 equidistant test points. 
The measurement range is defined via [SENSe:]POWer:TIME:DELay and [SENSe:]POWer:TIME:SPAN. 
Note:  The number of test points may be reduced at very narrow measurement ranges.