EasyManua.ls Logo

Rohde & Schwarz CMU 300 - Path Configuration - Subsystem Multitone:af1 Channel:filter

Rohde & Schwarz CMU 300
418 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CMU Audio Generator and Analyzer (with Option CMU-B41)
1100.4903.12 6.83 E-10
DEFault:MULTitone:AF1Channel:TDEFinition <Enable> Default Settings
DEFault:MULTitone:AF2Channel:TDEFinition <Enable>
<Enable>
Description of parameters Def. value Def. unit FW vers.
ON |
OFF
The parameters are set to their default values
Some or all parameters differ from the default values
ON
– V3.0
Description of command
If used as a setting command with the parameter ON this command sets all parameters of the subsystem to their
default values (the setting OFF results in an error message).
If used as a query the command returns whether all parameters are set to their default values (ON) or not (OFF).
DEFault:MULTitone:TDEFinition <Enable> Default Settings
<Enable>
Description of parameters Def. value Def. unit FW vers.
ON |
OFF
The parameters are set to their default values
Some or all parameters differ from the default values
ON
– V3.0
Description of command
If used as a setting command with the parameter ON this command sets all parameters of the subsystem (includ-
ing AF1Channel and AF2Channel) to their default values (the setting OFF results in an error message).
If used as a query the command returns whether all parameters are set to their default values (ON) or not (OFF).
Path Configuration – Subsystem MULTitone:AF1Channel:FILTer
The subsystem MULTitone:AF1Channel:FILTer configures the voice-processing equipment used for
the Multitone measurement. The subsystem corresponds to the Filters tab of the popup menu Multitone
Configuration.

Table of Contents

Related product manuals