CMU  RF Non Signalling Measurements 
1100.4903.12 2.9  E-9 
Additional Information...    Alternative Settings  
and Measurements 
... on Step 1
 
 
 
" Menu Select menu 
The Menu Select menu shows all function groups installed on 
your CMU. If a function group is selected the available test 
modes and measurement menus are indicated. Function 
groups representing digital network tests (such as 
GSM400/850/900/1800/1900-MS) are generally subdivided in 
the two test modes Non Signalling and Signalling, each con-
taining a number of measurement menus. 
The RF function group is available on any CMU regardless of 
the software options purchased. It comprises the three 
measurement menus Analyzer/Generator, Power and Spec-
trum. All three measurement menus are directly accessible 
from the Menu Select menu. 
 
 
&  Chapter 4 
For digital network tests refer to 
the relevant operating manuals. 
e.g. 
GSM400/850/900/1800/1900-MS.
... on Step 2
 
 
 
# Analyzer/Generator menu 
The Analyzer/Generator menu contains several softkeys to  
•  Control the RF signals received and analyzed (Analyzer 
Level, Analyzer Settings)  
•  Control the RF signals generated (Generator)  
Defining a level and frequency via the Generator softkey and 
the associated hotkeys implies that a continuous signal (CW) 
with this level and frequency is generated.  
The RF Max. Level defined via Analyzer Level, however, de-
notes the maximum input power which can be measured. 
This is identical with the upper edge of the Power diagram 
(see below). The permissible range of Max. Level depends 
on the input connector and external attenuation used (see 
section Analyzer Settings in Chapter 4). 
Defining a (center) Analyzer Settings – Frequency implies 
that only signals around this frequency are analyzed.  
The Analyzer Settings – RBW hotkey defines the resolution 
bandwidth of the analyzer. 
 
 
&
 Chapter 4 
The  Generator softkey provides 
the most important settings of the 
Generator tab of the Connection 
Control menu.  
The Analyzer Level and Analyzer 
Settings  softkeys correspond to 
the  Analyzer  tab of the Connec-
tion Control menu.  
 
$ The status of the Analyzer Power measurement is shown in 
the corresponding softkey. For ongoing measurements, the 
result in the Power output field is constantly updated. 
At present no input signal is available so that the Power out-
put field shows an invalid result “– – –“ 
Once the softkey is selected, the Analyzer Power measure-
ment can be switched off and on by means of the ON/OFF 
key. In contrast, the CONT/HALT toggle key halts the meas-
urement after the next valid result has been obtained.  
 
 
&
  Ch. 4 and Ch. 5 
The options for the measurement 
status are ON, OFF, or HLT. The 
HLT state is reached after the 
end of a single shot measurement 
(see the section about measure-
ment control in Chapter 5).