Remote Commands
R&S
®
FSW
721User Manual 1173.9411.02 ─ 19
Suffix:
<gap>
.
1 | 2
Gap channel number
Parameters:
<Spacing> numeric value in Hz
*RST: 2.5 MHz
Usage: SCPI confirmed
Manual operation: See "Gap (CACLR) Channel Spacings" on page 171
11.5.3.8 Performing an ACLR Measurement
The following commands are required to perform an ACLR measurement:
CALC:MARK:FUNC:POW:SEL ACP, see CALCulate<n>:MARKer<m>:FUNCtion:
POWer:SELect on page 698
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe] on page 698
INITiate[:IMMediate] on page 694, see chapter 11.5.1, "Performing Measure-
ments", on page 692
11.5.3.9 Analyzing Measurement Results
The following commands analyze and retrieve measurement results for ACLR meas-
urements.
Useful commands for channel power measurements described elsewhere
CALCulate:MARKer:FUNCtion:POWer<sb>:RESult? on page 696
TRACe<n>[:DATA] on page 922
Remote commands exclusive to channel power measurements
CALCulate<n>:MARKer<m>:FUNCtion:POWer:RESult:PHZ.............................................. 721
[SENSe:]POWer:ACHannel:MODE..................................................................................722
CALCulate<n>:MARKer<m>:FUNCtion:POWer:RESult:PHZ <State>
This command selects the way the R&S FSW returns results for power measurements.
You can query results with CALCulate:MARKer:FUNCtion:POWer<sb>:RESult?.
Parameters:
<State> ON | OFF
ON
Channel power density in dBm/Hz
OFF
Channel power in dBm
*RST: OFF