Remote Commands
R&S
®
FSW
749User Manual 1173.9411.02 ─ 19
Return values:
<RefRange> Number of the current reference range.
Range: 1 to 30
Usage: Query only
[SENSe:]ESPectrum<sb>:RTYPe <Type>
This command defines the type of the power reference.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
Parameters:
<Type> PEAK
Measures the highest peak within the reference range.
CPOWer
Measures the channel power within the reference range (integral
bandwidth method).
*RST: CPOWer
Manual operation: See "Power Reference Type" on page 217
11.5.6.6 Configuring the Power Classes
The following commands define the power classes for SEM measurements.
CALCulate:LIMit:ESPectrum<sb>:LIMits...........................................................................749
CALCulate:LIMit:ESPectrum<sb>:MODE..........................................................................750
CALCulate:LIMit:ESPectrum<sb>:VALue..........................................................................751
CALCulate:LIMit:ESPectrum<sb>:PCLass:COUNt.............................................................751
CALCulate:LIMit:ESPectrum<sb>:PCLass<class>[:EXCLusive].......................................... 752
CALCulate:LIMit:ESPectrum<sb>:PCLass<class>:LIMit[:STATe]........................................ 752
CALCulate:LIMit:ESPectrum<sb>:PCLass<class>:MAXimum.............................................753
CALCulate:LIMit:ESPectrum<sb>:PCLass<class>:MINimum..............................................753
CALCulate:LIMit:ESPectrum<sb>:LIMits <Max1>,<Max2>,<Max3>
This command sets or queries up to 4 power classes in one step. You can only define
values for the number of power classes defined by CALCulate:LIMit:
ESPectrum<sb>:PCLass:COUNt on page 751.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
Configuring and Performing Measurements