Remote Commands
R&S
®
FSW
774User Manual 1173.9411.02 ─ 19
Remote commands exclusive to spurious emission measurements
CALCulate<n>:PSEarch|PEAKsearch:AUTO.................................................................... 774
CALCulate<n>:PSEarch|PEAKsearch:MARGin.................................................................774
CALCulate<n>:PSEarch|PEAKsearch:PSHow.................................................................. 774
CALCulate<n>:PSEarch|PEAKsearch:SUBRanges........................................................... 774
CALCulate<n>:PSEarch|PEAKsearch:AUTO <State>
This command turns the list evaluation on and off.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
CALC:PSE:AUTO OFF
Deactivates the list evaluation.
CALCulate<n>:
PSEarch|PEAKsearch:MARGin <Threshold>
This command defines the threshold of the list evaluation.
Parameters:
<Margin> Range: -200 to 200
*RST: 200
Default unit: dB
Example:
CALC:PSE:MARG 100
Sets the threshold to 100 dB.
CALCulate<n>:PSEarch|PEAKsearch:PSHow
This command turns the peak labels in the diagram on and off.
Peak labels are blue squares.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:PSE:PSH ON
Marks all peaks with blue squares.
CALCulate<n>:PSEarch|PEAKsearch:SUBRanges <NumberPeaks>
This command defines the number of peaks included in the peak list.
After this number of peaks has been found, the R&S FSW stops the peak search and
continues the search in the next measurement range.
Parameters:
<NumberPeaks> Range: 1 to 50
*RST: 25