Remote Commands
R&S
®
FSW
775User Manual 1173.9411.02 ─ 19
Example:
CALC:PSE:SUBR 10
Sets 10 peaks per range to be stored in the list.
Manual operation: See "Peaks per Range" on page 248
11.5.7.4 Performing a Spurious Measurement
The following commands are required to perform a Spurious measurement:
SENS:SWE:MODE LIST, see [SENSe:]SWEep:MODE on page 731
INITiate[:IMMediate] on page 694, see chapter 11.5.1, "Performing Measure-
ments", on page 692
11.5.7.5 Retrieving and Saving Settings and Results
The following commands analyze and retrieve measurement results for Spurious
measurements.
Useful commands for spurious emission measurements described elsewhere
● CALCulate:LIMit<k>:FAIL? on page 987
● TRACe<n>[:DATA] on page 922
● TRACe<n>[:DATA]:MEMory? on page 924
● TRACe<n>[:DATA]:X? on page 924
11.5.7.6 Programming Example: Spurious Emissions Measurement
In the following example, the Spurious Emissions measurement is configured by defin-
ing ranges and parameters to create the following sweep list.
Configuring and Performing Measurements