Remote Commands
R&S
®
FSW
731User Manual 1173.9411.02 ─ 19
[SENSe:]ESPectrum<sb>:PRESet:STORe <Standard>
This command saves the current SEM measurement configuration.
Standard definitions are stored in an xml file. The default directory for SEM standards
is C:\r_s\instr\sem_std.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
Parameters:
<Standard> String containing the file name.
You can save the file in a subdirectory of the directory men-
tioned above. In that case, you have to include the relative path
to the file.
Manual operation: See "Save Standard" on page 223
11.5.6.2 Controlling the Measurement
The following commands control the measurement itself.
INITiate:ESPectrum .......................................................................................................731
[SENSe:]SWEep:MODE................................................................................................. 731
INITiate:ESPectrum
This command initiates a Spectrum Emission Mask measurement.
Usage:
Event
[SENSe:]SWEep:MODE <Mode>
This command selects the spurious emission and spectrum emission mask measure-
ments.
You can select other measurements with
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe]
Parameters:
<Mode> AUTO
Turns on basic spectrum measurements.
ESPectrum
Turns on spectrum emission mask measurements.
LIST
Turns on spurious emission measurements.
*RST: AUTO
Usage: SCPI confirmed
Manual operation: See "Spectrum Emission Mask" on page 113
See "Spurious Emissions" on page 113