Remote Commands
R&S
®
FSW
805User Manual 1173.9411.02 ─ 19
Parameters:
<Detector> OFF
no final measurement is performed
AVER
average detector
CAV
CISPR Average detector
CRMS
RMS Average detector
POS
maximum peak detector
QPE
quasipeak detector
*RST:
OFF
Manual operation: See "Final Test Detector" on page 301
11.5.13.3
Configuring the EMI Final Test
The commands required to configure the EMI final test are described here.
Useful commands for configuring EMI final tests described elsewhere:
[SENSe:]BANDwidth|BWIDth[:RESolution]:TYPE on page 837
[SENSe:]BANDwidth|BWIDth[:RESolution] on page 836
DISPlay[:WINDow<n>]:TRACe:X:SPACing on page 831
Remote commands exclusive to configuring EMI final tests
CALCulate:MARKer:FUNCtion:FMEasurement:PSEarch|PEAKsearch:AUTO...................... 805
CALCulate:MARKer:FUNCtion:FMEasurement:DWELl...................................................... 805
CALCulate:MARKer:FUNCtion:FMEasurement:PSEarch|PEAKsearch:AUTO
<State>
This command turns an automatic peak search on and off. If enabled, the search is
performed automatically after each sweep.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 0
Example:
CALC:MARK:FUNC:FME:PEAK:AUTO ON
Activates automatic marker peak search after each sweep.
Manual operation: See "Automatic Peak Search" on page 303
CALCulate:MARKer:FUNCtion:FMEasurement:DWELl <Time>
This command defines the dwell time during the final measurement.
Configuring and Performing Measurements