Remote Commands
R&S
®
FSW
985User Manual 1173.9411.02 ─ 19
Managing Limit Lines
Useful commands for managing limit lines described in the R&S FSW User Manual:
●
MMEM:SEL[:ITEM]:LIN:ALL
●
MMEM:STOR:TYPE
●
MMEM:LOAD:TYPE
Remote commands exclusive to managing limit lines:
CALCulate:LIMit:ACTive?............................................................................................... 985
CALCulate:LIMit<k>:COPY............................................................................................. 985
CALCulate:LIMit<k>:DELete............................................................................................985
CALCulate:LIMit<k>:STATe............................................................................................ 986
CALCulate:LIMit<k>:TRACe<t>:CHECk........................................................................... 986
CALCulate:LIMit:ACTive?
This command queries the names of all active limit lines.
Return values:
<LimitLines> String containing the names of all active limit lines in alphabeti-
cal order.
Example:
CALC:LIM:ACT?
Queries the names of all active limit lines.
Usage: Query only
Manual operation: See "Visibility" on page 530
CALCulate:LIMit<k>:COPY <Line>
This command copies a limit line.
Parameters:
<Line> 1 to 8
number of the new limit line
<name>
String containing the name of the limit line.
Example:
CALC:LIM1:COPY 2
Copies limit line 1 to line 2.
CALC:LIM1:COPY 'FM2'
Copies limit line 1 to a new line named FM2.
Manual operation: See "Copy Line" on page 531
CALCulate:LIMit<k>:DELete
This command deletes a limit line.
Usage:
Event