EasyManua.ls Logo

Rohde & Schwarz Scope Rider RTH1004 - 2.3 Input isolation

Rohde & Schwarz Scope Rider RTH1004
53 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preface
R&S
®
Scope Rider RTH
18Getting Started 1326.1561.02 ─ 10
IP rating
IP51 in line with IEC 60529
Safety compliance
IEC / EN / DIN EN 61010-1
IEC / EN / DIN EN 61010-2-030
IEC / EN / DIN EN 61010-2-033 (R&S RTH1002)
UL / CSA 61010-1
UL / CSA 61010-2-030
UL / CSA 61010-2-033 (R&S RTH1002)
EMC compliance
RF emission
In line with CISPR 11 / EN 55011 group 1 class A (for a shielded test
setup)
Complies with the emission requirements stipulated by EN 55011, EN
61326-1 and EN 61326-2-1 class A, making the instrument suitable for use
in industrial environments
Immunity: in line with IEC / EN 61326-1 table 2, immunity test requirements
for industrial environments
2.3 Input isolation
The instrument has independently floating isolated inputs. Each input channel has
its own signal input and its own reference input. Each input channel is electrically
isolated from the other input channels. Therefore, each reference of the used
inputs must be connected to a reference voltage. Furthermore, input channels are
electrically isolated from the communication ports and the power adapter input.
Input isolation

Other manuals for Rohde & Schwarz Scope Rider RTH1004

Related product manuals