Preface
R&S
®
Scope Rider RTH
19Getting Started 1326.1561.02 ─ 10
Figure 2-1: Isolation scheme of the R&S
RTH
The input isolation has several advantages:
●
You can measure independently floating signals simultaneously.
●
The risk of causing a short circuit while measuring multiple signals is reduced
substantially.
●
When measuring signals with different grounds, the induced ground currents
are kept to a minimum.
2.4 Documentation overview
The user documentation for the R&S RTH consists of the following parts:
●
Instrument Help
The instrument help is part of the instrument's firmware. It offers quick, con-
text-sensitive access to the complete information directly on the instrument.
●
Safety Instructions (multilingual)
Provides safety information in many languages. The printed document is
delivered with the product.
Documentation overview