Operating the instrument
R&S
®
Scope Rider RTH
42Getting Started 1326.1561.02 ─ 10
The [SHIFT] button toggles the wheel focus in the keypad. If the focus is
on the entry field, turning the wheel changes the value. If the focus is in
the lower part, the wheel selects numbers and unit.
5.3.3 Using front panel keys
For an overview of the front panel keys, see Figure 4-2.
Key Press shortly Press and hold
[AUTOSET] analyses the active channels,
adjusts the instrument settings, and displays
stable waveforms.
[PRESET] sets the instrument to the default
factory state.
[MEAS] starts or stops the last configured
automatic measurements.
Opens or closes the "Meas"
dialog to configure the mea-
surements.
[ZOOM] enables or disables the zoom with
the last configuration.
If the zoom is on but not in focus, pressing
the key focuses the zoom.
Opens or closes the "Zoom"
dialog to configure the zoom
scale and position.
[CURSOR] starts or stops the last configured
cursor measurement.
If the cursor is on but not in focus, pressing
the key sets the focus to the first cursor line.
Opens or closes the "Cursor"
dialog to configure the mea-
surement.
[MATH] switches the math waveform on or
off.
Opens or closes the "Math"
dialog to configure the math
waveform.
Requires logic analyzer option R&S RTH-B1
(MSO).
The effect depends on the state of digital
channels:
If the all digital channels are off, the key
switches them on and sets the focus.
If the digital channels are on but not in focus,
the key sets the focus.
If the focus is on digital channels, the key
switches them off.
Opens or closes the "Logic"
dialog to configure digital
channels.
Activates or deactivates the serial bus. Key
function requires at least one serial triggering
and decoding option. Available options are
listed in the specifications document.
Opens or closes the "Bus"
dialog to configure serial pro-
tocols.
Accessing the functionality