50
Details of the Auto Operation Mode
• Embedding Module
When the start time arrives, the instrument is automatically activated to
make sure parafn melts sufciently.
When the end time arrives, the instrument automatically pauses.
(Active)
4 hours
Preparing OperatingPaused
0 AM 0 AM 0 AM
Day 1 Day 2
Start
time 1
End
time 1
Start
time 2
End
time 2
(*) [Start time 1] and [End time 1] represent the start time and end time on day 1.
[Start time 2] and [End time 2] represent the start time and end time on day 2.
Paused
Instrument
state
TEC 6-EM
(Active)
Preparing Operating
Paused
4 hours
• Example of special operation in auto operation
* If a time earlier than start time 1 is set to end time 1, operation will
continue after the 24:00 hour (0 hour) and end at end time 1 on the
following day. Depending on the combination of the start time and the
end time on the following day, operation may continue after end time 1.
Preparing Operating
0 AM 0 AM 0 AM
Day 1 Following day
Start
time 1
End
time 1
End
time 1
Start
time 2
End
time 2
Example 1 [End time 1] is at least 4 hours earlier than the start time on the following day ([Start time 2]):
> Operation will end at [End time 1].
Paused Preparing Operating Paused
4 hours24 hours
End