52
The “Extend” button is touched while the instrument is paused or
preparing:
The auto operation will continue until the end time of the next day if the
auto operation of the day has already passed the end time.
Operating Preparing OperatingPaused
0 AM 0 AM 0 AM
Day 1 Day 2
Display of the
"Extend" button
Paused
Inactive Active Inactive
Instrument state
End
End
time 1
Start
Time 2
End
time 2
“Extend” button
(Nearby
Set Temp)
If the auto operation of the day has not passed the end time, auto
operation of the day will resume. The operation will end at the end time of
the day.
NOTE:
Touch the “Extend” button again to extend the operation
continuously till the end time for the next day and later.
End
Operating OperatingPaused
Resumed
operation
PausedPreparing
0 AM 0 AM
Day 1 Day 2
Display of the
"Extend" button
Prep.
(4 hrs)
Instrument state
Start
Time 1
End
time 1
End
Paused
End
time 2
Start
Time 2
“Extend” button“Stop” button
Inactive
Although the “Extend” button is pressed, the display does not change because it is
only resuming from pause. When the “Extend” button is pressed again after resuming
from pause, the operation will be performed until the end time of the next operation
NOTE:
In the following cases, the extension mode will be switched OFF
from ON and the “Extend” button turns inactive.
When the “Extend” button is touched again while in the extension
mode.