3. Explanation of Operations
57
6. Add approx. 80 g into the warming chamber, and then insert the large tray.
This parafn is used to keep the distribution of tray temperatures uniform.
Add enough parafn to uniformly cover the bottom of the tray.
NOTES:
In the figure above, the right warming chamber is set up. The left
warming chamber can be set up in the same manner.
When the optional small tray is used, the two small trays need to
be placed in one warming chamber. Placing only the one small
tray may cause thermal conduction to slow.
7. It is posiible to use parafn in the warming chamber tary to soak cassettes
containing tissues.
Add parafn to the graduation line marked on the warming chamber tray.
The tray containing parafn is called the “Tissue Tray.”
り線
まで
体ト
CAUTION:
Do not overfill the tray. Doing so may cause paraffin to spill
out, resulting in instrument failure and/or accident.
Add parafn
to the
marked line.
Tissue tray