EasyManua.ls Logo

Sakura Tissue-Tek TEC 6 - Page 58

Sakura Tissue-Tek TEC 6
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
58
8. Put the base molds into the warming chamber in which the tissue tray is
not stored.
9. Put approx. 2 g of parafn into the forceps holder, and heat the forceps to
be used in the holder.
10. Put tampers on the hot plate and heat the tampers.
タン
使用際、固
ィン
ット
しま
ット
11. The embedding process can be started once the paraffin has melted in
the paraffin chamber and warming chamber and the paraffin chamber
temperature indicators change from a blinking light to a steady light.
Tampers
Heat the tampers on the hot plate so that
solid parafn will not attach to the tampers
when they are used.
Hot plate.

Table of Contents

Other manuals for Sakura Tissue-Tek TEC 6