60
3. Move one cassette from the tissue tray onto the hot plate.
4. Remove the lid from the cassette.
5. Remove a base mold of a size suitable for the tissue from the warming
chamber.
6. Press the nger plate with forceps or a nger to dispense a small amount
of parafn onto the base mold.
NOTE:
To make the area around the finger plate brighter, touch the “Light”
button on the Operation screen to turn on the work light.
If the optional foot pedal is connected, paraffin
can be dispensed by stepping on the foot pedal,
instead of pressing the finger plate.
ット
Hot plate
ィン
底か 1~2
mm
ィン
Finger plate
1 to 2 mm of parafn
from bottom

Table of Contents