3. Explanation of Operations
61
7. Place the tissue into the base mold and slightly press to the bottom by
forceps.
8. Cool the bottom of the base mold with the cold spot, and push the tissue
with a tamper to x its position. (Cool the bottom of the base mold until a
thin layer of solidied parafn is formed.)
9. Place the cassette on the base mold with the tissue side up.
10. Press the nger plate and add parafn to around 80% of the depth of the
cassette.
ット
タン
Tamper
Cold spot
の深の、
8 分目程度
Add parafn to around 80% of the
depth of the cassette.