62
11. Place the base mold onto the cooling plate of the Cyro Module and cool
the mold.
NOTE:
To cool the paraffin block completely, place the base mold so that
its bottom contacts with the cooling plate.
12. Wait for several minutes until the paraffin has solidified, remove the
parafn block from the base mold.
13. Clean the tamper and the interior surface of the base mold using a lint-
free paper.
If the interior surface of the base mold is dirty, the parafn block may not
be removed easily in the next process.
If there are still tissues to be embedded, repeat the steps from 3.
ィン
Completed parafn block
包埋皿の内表面(赤色の部分)
Interior surface of a base mold
(red area)