5. Troubleshooting
83
Error
code
Overview
Displayed
message
Description
Instrument operation
when the error occurs
Solution
E040 Low battery
caution
Battery voltage
has dropped
The voltage of the clock
battery dropped below
the specied value.
Operation continues. Contact the Sakura
instrument distributor or
representative and replace
the battery within one
month.
E041 Clock reset The clock was
reset.
Unable to restore time
data. Button battery is
dead.
If the instrument was
operating in the auto
mode, operation
pauses. If the instrument
was operating in the
manual mode, operation
continues.
Contact the Sakura
instrument distributor or
representative and replace
the battery.
E042 Invalid clock The time data is
incorrect
The clock generated an
error. The battery case
came off.
Operation continues. Check if the date and time
displayed on the screen is
correct. If the date and time
is not correct, set the correct
date and time. If the problem
persists, contact the Sakura
instrument distributor or
representative and replace
the battery.
E050 Momentary
power failure
Momentary
power outage
A momentary power
failure occurred.
Operation continues
from the Operation
screen. If the light was
on when the power
was cut off, operation
continues with the light
turned off.
Check the molten state of
paraffin, and then resume
using the instrument.
E100 Abnormal
tempera-ture
sensor
Temperature
sensor error
Signals cannot be
received from the
temperature sensor.
Operation aborts. Turn off the power to the
instrument and contact
the Sakura instrument
distributor or representative.
E101
E102 An abnormal signal
was received from the
temperature sensor.
Operation aborts. Turn off the power to the
instrument. When parafn in
the problem location starts
solidifying (below 60ºC, as
a guide), turn on the power
again. (The power should be
off for at least 10 seconds.)
If the problem persists,
turn off the power to the
instrument and contact
the Sakura instrument
distributor or representative.
(Check to see if any
equipment that generates
strong electromagnetic
waves is nearby.)
E103 Abnormally
high
tempera-ture
Abnormally high
tempera-ture
The temperature of each
location exceeds the
allowable value.
(The temperature is
higher than the E010
level.)
Operation aborts. Turn off the power to the
instrument. When parafn in
the problem location starts
solidifying (below 60ºC, as
a guide), turn on the power
again. (The power should be
off for at least 10 seconds.)
If the problem persists,
turn off the power to the
instrument and contact
the Sakura instrument
distributor or representative.
(Check to see if any
equipment that generates
strong electromagnetic
waves is nearby.)
High temperature molten
parafn was added.
E901 Abnormal
nonvolatile
memory
Memory error
during data
update
Data could not be saved
because the nonvolatile
memory operated
abnormally.
Operation aborts. Operation aborts. Turn off
the power to the instrument
and contact the Sakura
instrument distributor or
representative.
E902
E903
E904