84
Descriptions of Additional Codes
Additional code Description (Applicable location)
01 Cooling plate (Cryo Module)
02 Parafn chamber (Embedding Module)
03 Left side warming chamber (Embedding Module)
04 Right side warming chamber (Embedding Module)
05 Left side hot plate (Embedding Module)
06 Right side hot plate (Embedding Module)
07 Parafn line (Embedding Module)
08 Dispensing solenoid valve (Embedding Module)
09 Dispenser (Embedding Module)
10 Forceps holder (Embedding Module)
11 Hot plate heater
21 Set value data
22 Operation status data
23 Dispensing log data
24 Error log data
25 Temperature record data
Buzzer Notication
This instrument noties the following events with a buzzer sound.
Type of notication Notication sound Condition for notication How to stop
Startup notication Short beep The display has started after the
power was turned on.
(Approx. 15 seconds later)
Stops automatically.
Start notication Short beep The TEC 6-EM is operating. Stops automatically.
End notication Short beep Auto operation has ended. Stops automatically.
Dispensing notication 3 short beeps An attempt was made to perform
dispensing operation when dispensing
was not possible.
Stops automatically.
Parafn solidied
notication
15-second continuous
beep
The parafn solidication process
was complete in the cleaning mode.
Stops automatically or by
toughing the “
” (stop) button.
Parafn heated
notication
Intermittent short beeps
(for one minute)
The parafn melting process was
complete in the cleaning mode.
Stops automatically or by
toughing the “
” (stop) button.
Error notication
(Caution)
One-minute continuous
beep
An caution-level error occurred Stops automatically or by
toughing the “
” (stop) button.
Error notication
(Warning)
Continuous beep An warning-level error occurred.
Touch the “
” (stop) button.
NOTE:
The buzzer notification cannot be disabled. The volume of the
buzzer is fixed.