Operating Manual Biowelder
®
 Total Containment  39
  Appendix 
9  Appendix
9.1  Copyright
Parts of the device are patented. No part this device may be reproduced, 
distributed, translated or transferred, in any form or through any means, 
electronically or  mechanically, including photocopying, recording or saving 
to an information storage and retrieval system, without the previous written 
approval of Sartorius Stedim  Biotech GmbH.
9.2  Disposal
The manufacturer disposes of the Biowelder
®
 TC.
The device must be decontaminated before being sent back to Sartorius 
Stedim Biotech.
9.3  Technical data
Type designation Biowelder
®
 TC, BWTC
Power connection 100 VAC – 240 VAC
Input frequency 50/60 Hz
Power input 300 VA
In and out connections Device plug C14 max. 250VAC
Ethernet jack type RJ45
Power connection of fuse 2 × 3.15 A T (Type FST)
Battery CR2032
Operating temperature +5°C – +40°C *
Place of use Indoor (Laboratory)
Transient overvoltage Overvoltage category II
Pollution degree 2
Altitude up to 2000 m
Humidity 80% up to 31°C, linearly diminishing to 
50% relative humidity at 40°C, not 
condensing
Degree of protection IP20
Weight 16.4 kg
External size (L × W × H) 555 mm × 261 mm × 269 mm
Power cord According to local regulations  
minimum 3 × AWG18 or 3 × 0.75 mm
2
 
minimum local mains supply voltage
 *  The device is programmed with standard parameter sets for welding C-Flex
®
 374, 
AdvantaFlex
®
, PharMed
®
 BPT and SaniPure
™
 BDF
™
.  
These parameter sets have been validated at room temperature.